Aortic prosthetic valve calculator: EOA, DVI and PPM
Review effective orifice area, Doppler velocity index and prosthesis–patient mismatch inputs for the aortic prosthesis tool.
Method
EOA = LVOT area × LVOT VTI ÷ prosthesis VTI; DVI = LVOT VTI ÷ prosthesis VTI; iEOA = EOA ÷ BSA
Use the measurement units shown below. Some fields apply only to a selected method or condition; follow the interactive tool’s method choices. The interactive tool keeps missing measurements incomplete and provides its method-specific results and reference context.
Measurements and acquisition
LVOT VTI cm
Pulsed-wave Doppler
- Place the sample volume at the specified annulus or outflow tract.
- Align the Doppler beam with flow and record a clean envelope.
- Trace the modal velocity envelope and use the reported VTI.
Prosthesis VTI cm
Continuous-wave Doppler trace
- Use the same aligned CW recording as the peak velocity.
- Trace the outer edge of the dense envelope for the relevant phase.
- Use the machine-reported VTI in centimeters.
LVOT diameter cm
2D echo · inner-edge measurement · Optional
- Use the specified view and cardiac phase.
- Align the calipers perpendicular to the structure’s long axis.
- Measure inner edge to inner edge and avoid oblique foreshortening.
Body surface area m²
Record the units, date and measurement source · Optional
- Use measured height and weight where feasible. Record the units, date and whether a value was reported or estimated.
- Fluid status and weight changes affect BSA. Document any substituted weight; do not silently replace measured weight with ideal or dry weight.
- Use the unit named by the field. BSA is in m²; height and weight can be converted in the Body surface area calculator. Keep the formula with serial indexed results.
Body mass index kg/m²
Record the units, date and measurement source · Optional
- Use measured height and weight where feasible. Record the units, date and whether a value was reported or estimated.
- Fluid status and weight changes affect BSA. Document any substituted weight; do not silently replace measured weight with ideal or dry weight.
- Use the unit named by the field. BSA is in m²; height and weight can be converted in the Body surface area calculator. Keep the formula with serial indexed results.
Acceleration time ms
Spectral Doppler timing · Optional
- Use a sweep speed that clearly shows the interval boundaries.
- Measure from the specified onset to offset.
- Average representative beats using consistent rhythm handling.
Ejection time ms
Spectral Doppler timing · Optional
- Use a sweep speed that clearly shows the interval boundaries.
- Measure from the specified onset to offset.
- Average representative beats using consistent rhythm handling.
Sources and method context
American Society of Echocardiography · 2024 · Guidelines for the Evaluation of Prosthetic Valve Function With Cardiovascular Imaging · EOA and DVI pp. 7–9 · Table 5 JASE p. 19 · Table 7 p. 20
- EOA and DVI calculation · Section IC(iii)(b–c) · PDF pp. 7–9
Continuity-derived effective orifice area, LVOT sampling, and Doppler velocity index.
- Acceleration indices · Table 5 · JASE p. 19
Acceleration time and AT/ET ranges for normal, possible, and significant stenosis patterns.
- BMI-specific aortic PPM · Table 7 · JASE p. 20
Indexed EOA thresholds for BMI below 30 and BMI at least 30 kg/m².
Catalogue review status: Exact guideline locator mapped · clinical sign-off pending
These pages describe the educational tool. Interpret measurements with the complete clinical context and the method’s acquisition limitations.
Keep exploring
- Parasternal long axis acquisition guide · Open in Probe Lab
- Apical five chamber acquisition guide · Open in Probe Lab
- Apical three chamber (apical long axis) acquisition guide · Open in Probe Lab
- Surgical aortic prosthesis · obstruction calculator
- Aortic prosthesis · serial comparison calculator
- Mitral prosthetic valve calculator: EOA, DVI and PPM
- Mitral prosthesis · obstruction calculator