Qp/Qs calculator: pulmonary to systemic flow ratio
Review the Doppler stroke-volume measurements used for pulmonary and systemic flow in the Qp/Qs tool.
Method
Qp/Qs = (RVOT CSA × RVOT VTI) ÷ (LVOT CSA × LVOT VTI)
Use the measurement units shown below. Some fields apply only to a selected method or condition; follow the interactive tool’s method choices. The interactive tool keeps missing measurements incomplete and provides its method-specific results and reference context.
Measurements and acquisition
RVOT maximal-systolic diameter at PW site cm
2D echo · inner-edge measurement
- Use the specified view and cardiac phase.
- Align the calipers perpendicular to the structure’s long axis.
- Measure inner edge to inner edge and avoid oblique foreshortening.
RVOT VTI cm
Pulsed-wave Doppler
- Place the sample volume at the specified annulus or outflow tract.
- Align the Doppler beam with flow and record a clean envelope.
- Trace the modal velocity envelope and use the reported VTI.
LVOT maximal-systolic diameter at PW site cm
2D echo · inner-edge measurement
- Use the specified view and cardiac phase.
- Align the calipers perpendicular to the structure’s long axis.
- Measure inner edge to inner edge and avoid oblique foreshortening.
LVOT VTI cm
Pulsed-wave Doppler
- Place the sample volume at the specified annulus or outflow tract.
- Align the Doppler beam with flow and record a clean envelope.
- Trace the modal velocity envelope and use the reported VTI.
Sources and method context
ASE / SCAI · 2015 · Echocardiographic assessment of atrial septal defect and patent foramen ovale · Pulsed Doppler shunt-flow quantification · pp. 929–930
- Pulmonary-to-systemic flow ratio · pp. 929–930 · Doppler shunt quantification
Each outflow stroke volume is its cross-sectional area times matched systolic VTI. Qp/Qs is RV outflow stroke volume divided by LV outflow stroke volume. Circular-area assumptions and semilunar regurgitation limit the estimate.
- Adult shunt context · AHA/ACC 2025 · ACHD guideline · §4.1 / Table 25
A Qp/Qs of at least 1.5 is interpreted with chamber loading, PVR and the full congenital assessment. The app’s greater-than-2 interval is a flow-reporting descriptor, not a universal defect severity or closure rule.
Catalogue review status: Exact guideline locator mapped · clinical sign-off pending
These pages describe the educational tool. Interpret measurements with the complete clinical context and the method’s acquisition limitations.
Keep exploring
- Parasternal long axis acquisition guide · Open in Probe Lab
- Apical five chamber acquisition guide · Open in Probe Lab
- RV outflow acquisition guide · Open in Probe Lab
- Parasternal short axis, aortic valve level acquisition guide · Open in Probe Lab
- Atrial septal defect: shunt and Doppler calculator
- VSD · size & right-heart pressure calculator
- PDA · shunt & PA pressure calculator
- Coarctation · anatomy & gradients calculator