Aortic regurgitation PISA calculator
Review the PISA inputs and Doppler acquisition guidance used to quantify aortic regurgitation.
Method
EROA = 2πr²Vₐ ÷ Vmax; RVol = EROA × VTI
Use the measurement units shown below. Some fields apply only to a selected method or condition; follow the interactive tool’s method choices. The interactive tool keeps missing measurements incomplete and provides its method-specific results and reference context.
Measurements and acquisition
PISA radius cm
Color Doppler · peak-flow frame
- Zoom on the valve and center the flow-convergence region.
- Shift the color baseline until the shell is as hemispheric as possible.
- Measure from the regurgitant orifice to the first aliasing boundary along the jet centerline.
Aliasing velocity cm/s
Shifted color scale · absolute cm/s
- Use the same color-Doppler frame used for the radius.
- Read the Nyquist limit after baseline shift, not the jet velocity.
- Enter the absolute displayed value in cm/s.
AR Vmax m/s
Continuous-wave Doppler
- Align the CW cursor as parallel as possible to the jet.
- Acquire a dense, complete envelope without truncation.
- Measure the highest absolute velocity on the selected envelope.
AR VTI cm
Continuous-wave Doppler trace
- Use the same aligned CW recording as the peak velocity.
- Trace the outer edge of the dense envelope for the relevant phase.
- Use the machine-reported VTI in centimeters.
LVOT stroke volume mL
Enter the measured value in the displayed unit
- Confirm the variable matches the selected formula.
- Use the unit displayed beside the field.
- Keep measurements from the same representative hemodynamic state.
Sources and method context
American Society of Echocardiography · 2017 · Recommendations for Noninvasive Evaluation of Native Valvular Regurgitation · AR PISA and flow-convergence method pp. 34–39 · Table 10 pp. 35–37 · Table 11 p. 38 · Figure 21 p. 39 · integrative pathway pp. 41–42
- Equation, phase & acquisition · AR section p. 34 · Table 10 pp. 35–37 · Figure 21 p. 39
Early-diastolic radius and peak AR velocity pairing, baseline shift, CW peak/VTI acquisition, and the EROA and regurgitant-volume equations.
- Quantitative grades · Table 11 · PDF p. 38
Four quantitative EROA, regurgitant-volume, and regurgitant-fraction bands, including the two moderate subgrades.
- Limitations & severity integration · AR section pp. 34–37 · integrative approach pp. 41–42
Far-field and calcification limitations, nonhemispheric or multiple-jet limitations, and integration with anatomy, vena contracta, aortic flow reversal, LV remodeling, hemodynamics, and data quality.
Catalogue review status: Exact guideline locator mapped · clinical sign-off pending
These pages describe the educational tool. Interpret measurements with the complete clinical context and the method’s acquisition limitations.
Keep exploring
- Parasternal long axis acquisition guide · Open in Probe Lab
- Apical five chamber acquisition guide · Open in Probe Lab
- Apical three chamber (apical long axis) acquisition guide · Open in Probe Lab
- Aortic regurgitation calculator: continuity method
- Mitral regurgitation PISA calculator: EROA and regurgitant volume
- Mitral regurgitation calculator: continuity method
- Tricuspid regurgitation calculator: PISA quantification